5.0 FUNCTION CHECK

Perform all steps in this procedure before using the table. For a complete definition of terms
used in this procedure, please refer to the Glossary of Terms in section 1.6 of the manual.
1. Turn the power switch off. Plug in the hand pendant cable connector into the hand
pendant receptacle on the control panel.
2. If the table is to be used under battery power, put the power switch in the on position and
wait for 20 seconds for the table to power up and run its self diagnostics. The battery
power lamp on the control panel should illuminate.
This symbol indicates an external ground stud that is required for use when the
AC power cable is not connected to a protective earth ground hospital grade AC outlet
in your operating room or facility.
To protect the patient, hospital staff and the table from possible electrical hazards, an
external ground wire connection is required between the external ground stud and
protective earth ground.
3. Battery operation check:
3.1 Observe table battery power status light in the power indicator section of the control
panel or on the hand pendant. A green light indicates the table is set for battery
power operation.
3.1.1 If green battery Ok light is on, the table is ready to operate normally.
3.1.2 If red battery charge light is on, the battery must be charged prior to using the
table. To charge the battery, make sure the power cord is plugged into a live
receptacle and turn on the power switch. This switch will illuminate indicating
that appropriate power is applied to the table. The table must remain plugged in
and switched on for a minimum of 3 hours to insure sufficient charging of the
battery to operate the table.
3.1.3 If red check battery light remains on after 3 hours, continue to charge battery for
up to 18 hours.
3.1.4 If green battery ok light does not illuminate after 18 hours, refer to section 6.5 of
this manual.
NOTE: The table may be used with AC power even if the battery status light is red, indicating
batteries need charging.
4. If the table is to be used under line power, plug power cord into an appropriate hospital
grade AC outlet and turn on the power switch. Note that the power switch illuminates
indicating the AC power on. The AC power lamps on the control panel and hand pendant
should illuminate.
5. Hand pendant check:
5.1 Press and hold HEIGHT UP button. Observe that the tabletop moves up.
5.2 Press and hold HEIGHT DOWN button. Observe that the tabletop moves down.
5.3 Press and hold REVERSE.TRENDELENBURG button. Foot end of tabletop becomes
lower than the head end
MIZUHOSI 2009