Sterilization:
NOTE: Mizuho OSI does not recommend the 6850-919, 6850-918 femoral hooks, left & right
or 6850-110 femoral hook support to be sterilized by flash or chemical sterilization.
Sterilization of the femoral hook(s) and femoral hook support is accomplished by steam. To
achieve a sterility assurance level of 10
parameters listed below:
Sterilization Method
Gravity Steam
NOTE: Other sterilization cycles may also be suitable. However, individuals or hospitals not
using the recommended methods are advised to validate any alternative methods using
current good laboratory practices (cGLP).
Storage:
1. Store the Mizuho OSI femoral hook(s) and femoral hook support with care to prevent
damage.
2. Keep stored instruments on carts or shelving in a storage area free of dust, insects,
chemical vapors and extreme changes in temperature and humidity.
Maintenance and repair of femoral hooks and femoral hook support:
Instruments returned to Mizuho OSI for repair require a certificate of disinfection
which testifies that each instrument has been thoroughly cleaned and disinfected.
Failure to supply a certificate of cleaning and disinfection will result in a cleaning
charge and delayed processing of your instrument repair.
1. Repair of the femoral hook(s) and femoral support hook by parties other than Mizuho OSI
will void the above warranty.
2. If the device requires repair or maintenance contact your local Mizuho OSI sales
representative or Mizuho OSI Customer Resource Group for repair or replacement.
Utilization:
Prior to the anterior approach total hip arthroplasty procedure, the appropriate right or left
femoral hook (6850-919 or 6850-918) and femoral hook support should be cleaned,
disinfected and sterilized according to the hospital's recommended guidelines. During the
surgical procedure, the surgeon will drape the femur lift and place the sterile, femoral hook
support and sterile, femoral hook in the desired position. The height of the femoral hook
support is raised and lowered by depressing the foot pedal.
MIZUHOSI 2009
-6
, Mizuho OSI recommends the sterilization
Temperature
Containment Type
270°F/134°C
WARNING:
31
Minimum Exposure
Wrapped
Time Range
10-25 Minutes
NW0508 Rev. D