Trendelenburg or reverse trendelenburg not functioning.
 Verify power cord is plugged into an appropriate AC receptacle and into the power
cord socket (located on table-base panel).
 Press and hold hand pendant trendelenburg or reverse trendelenburg control button.
Observe tabletop movement.
 Verify hand pendant is properly plugged into a hand pendant receptacle; observe
hand pendant control buttons illuminate green.
NOTE: Table power switch must be turned off before disconnecting or connecting
hand pendant.
 With table off plug hand pendant into the other hand pendant receptacle, turn power
switch on; observe hand pendant control buttons illuminate green.
NOTE: Table power switch must be turned off before disconnecting or connecting
hand pendant.
 Verify if other hand pendant control buttons are functional. If so, replace hand
pendant.
NOTE: Table power switch must be turned off before disconnecting or connecting
hand pendant.
 If on battery power, verify proper charge; observe battery status LED illuminates
green.
 Verify power switch is on.
 Call Mizuho OSI Customer Resource Group (800-777-4674) for parts or repair.
Left lateral roll or right lateral roll not functioning.
 Verify power cord is plugged into an appropriate AC receptacle and into the power
cord socket (located on table-base panel).
 Press and hold hand pendant left lateral roll or right lateral roll control button. Observe
tabletop movement.
 Verify hand pendant is properly plugged into a hand pendant receptacle; observe
hand pendant control buttons illuminate green.
NOTE: Table power switch must be turned off before disconnecting or connecting
hand pendant, then re-power on.
 Verify if other hand pendant control buttons are functional. If so, replace hand
pendant.
NOTE: Table power switch must be turned off before disconnecting or connecting
hand pendant, then re-power on.
 If on battery power, verify proper charge; observe battery status LED illuminates
green.
 Verify power switch is on.
 Call Mizuho OSI Customer Resource Group (800-777-4674) for parts or repair.