Plug the foot pedal into the port at the foot end of the table base. (The flange should be flush
or below surface of receptacle).
Foot pedal
port
Ensure the foot pedal and femur lift function correctly. On the control panel, (see figure 4)
verify the femur lift is connected by the corresponding LED illuminating when the appropriate
lift is selected. Depress the foot pedal, up side and observe the lift rise to its fullest height
and then depress down side and observe the lift lower to its lowest position. Complete this
process for both the right and left femur lifts.

8.4 Femoral Hook Support And Femoral Hook(s) Sterilization Guidelines

The femoral hook(s) (6850-919, 6850-918) and femoral hook support (6850-110) are
designed for use with only the Mizuho OSI 6850 PROfx
Mizuho OSI 6875 hana
Sparing (SITS) Anterior Approach Total Hip Arthroplasty. The hooks are imprinted with an L
and R to designate left and right.
Figure 23: The femur lift is in place for use with the femoral hook support 6850-110
Proper cleaning, handling and sterilization will ensure that the femoral hook(s) and femoral
hook support perform as intended.
MIZUHOSI 2009
Figure 22: Plugging in the foot pedal
®
Surgery Table. They are utilized during the Single Incision Tissue
®
Orthopedic Surgery Table or the
6850-918/919 Femoral
hook
6850-110 Hook
support
Lift
29
Flange
Below this
surface
NW0508 Rev. D