3.4 Hand Pendant/Control Panel Operation
®
The 6875 hana
via control panel built into power supply. The hand pendant is plugged into the receptacle
labeled hand pendant on the control panel (see figure 3). To activate a desired function,
press and hold the appropriate button on the hand pendant until the desired position is
achieved.
The hand pendant is equipped with a return to level button. This button functions to first level
the lateral roll, second level the trendelenburg, third re-level lateral roll and fourth brings the
table to its lowest position. A delay of 3 seconds is programmed after the leveling functions
are completed before returning to its lowest position. To achieve the final position, this
button must be held continuously until motion stops. The controller considers the table to be
level if it is within +/- 2 degrees of horizontal. If the return to level button is pressed while the
table is within this range, it may not move. If it is required to adjust the table to a position
closer to zero degrees, press the appropriate function button on the hand pendant until the
desired position is achieved.
Height Up
Trendelenburg
Left Lateral Roll
®
Figure 7: hana
When installing a hand pendant, turn the power switch off, then plug in hand pendant
connector and rotate locking collar clockwise until hand tight. To operate any of the
desired functions, turn the power switch on and wait for 20 seconds for the table to
power up and run its self diagnostics.
MIZUHOSI 2009
Surgery Table is controlled by means of a hand pendant, alternate control,
Height Down
Reverse
Trendelenburg
Right Lateral Roll
Surgery Table 6807-4 hand pendant
Height Up
Trendelenburg
Left Lateral Roll
Figure 8: Head end control panel
WARNING:
12
Height Down
Reverse
Trendelenburg
Right Lateral
Roll
NW0508 Rev. D