8.0 SET-UP OF 6875 hana
SURGERY TABLE
®
Instructions are provided for the following components to the 6875 hana
Surgery Table:
Leg spars 6875-350
Femur lift(s) and foot pedal assembly 6875-600, 6875-601, 6875-19
Femoral support and femoral hook(s) 6850-110.6850-919, 6850-918
Well leg support adaptor 6875-201
®
hana
Knee Flexion System™ for TKA 6875-230
8.1 Insertion And Removal Of Leg Spars – 6875-350 & 6875-360
Raise the height of the table to be able to easily access the spar mount. Insure the spar
locking knob is in the unlocked position by turning counterclockwise. Prior to mounting the
spar, ensure the spar ball joint handle is in the locked position. One person needs to support
the distal (foot) end of the spar holding the loop handle, while another person supports the
spar as it slides into the spar mount on the table. Slide the end of the spar forward until it is
fully engaged, and a defined click should be heard when spar post is inserted past the safety
latch. Lock in place by turning the spar locking knob clockwise until tight. Slight movement
of spar up and down while tightening knob will insure a tighter lock. While continuing to
support the distal (foot) end of the spar holding the loop handle, move the spar lock handle to
the unlocked position. Reposition the spar as needed and then return the spar ball joint
handle to the locked position.
Spar mount
Spar
Spar locking knob
Figure 13: Mounting the leg spar
Unlock
Lock
Safety latch
Figure 14: Locking and unlocking the leg spar
To remove the spar from the table, one person should support the distal (foot) end of the
spar with the loop handle and one person should turn the spar locking knob counterclockwise