8.14 Hip Pinning: Supine With Bilateral Skin Traction 6875 hana

Figure 36: Patient positioned for supine for hip pinning with bilateral skin traction
Instructions provided are for suggested set-ups. Final determination of patient positioning
and set-up is at the discretion of the surgeon and surgical team.
1. Ensure table floor locks are completely deployed and floor lock on is illuminated on
the hand pendant.
2. Attach radiolucent leg spars (6875-360 and 6875-350) to the table and lock in place.
Refer to section 8.1.
3. Place pelvic pad (6875-7129) onto table.
4. Attach lower leg support (6850-170) to non-operative leg spar using accessory clamp
(5855-61) and lock in place.
5. Position patient safety strap (5840-44) in place.
6. Adjust table height to appropriate position to transfer patient from patient stretcher.
7. Attach appropriate traction boots (6850-485) small, or (6850-486) large to patient's
feet while on the patient stretcher.
8. Transfer patient to the table in supine position, insert well padded perineal post and
pad (6850-413) move the patient to establish firm contact with perineal post and
secure with patient safety strap (5840-44).
9. Attach traction boots to leg spar traction mounts and lock in place.
10. Install arm boards with pad (6875-292) to each side of the table and position patient's
arms out to the side, or attach arm boards with pad (6875-292) to the non operative
side of the table and optional cross arm support (5857) using clark socket (5393) to
the non-operative side of the table and secure the patient's arms.
11. Trial position the C-Arm.
12. Install the head end drape rod assembly (6875-747) at the head end of the table and
the foot end drape rod assembly (6875-750) on the appropriate leg spar at the foot
end of the table with hole on the outside. Adjust to the appropriate height and lock.
MIZUHOSI 2009