8.13 Hip Pinning: Supine With Unilateral Skin Traction 6875 hana
Figure 35: Patient positioned for Right Hip Pinning, supine position with unilateral skin traction and well leg
supported in leg holder
Instructions provided are for suggested set-ups. Final determination of patient positioning
and set-up is at the discretion of the surgeon and surgical team.
1. Ensure all (4) caster floor locks are in the locked position.
2. Attach appropriate radiolucent leg spar (6875-360) right or (6875-350) left to the table
and lock in place. Refer to section 8.1.
3. Place pelvic pad (6875-7129) onto table.
4. Attach optional patient transfer board (6875-742) by clamping the device to the leg
spar on the side of the table from which the patient will be transferred to the table.
Refer to section 8.6.
5. Attach well leg support adaptor (6875-200) refer to section 8.5 into the leg spar mount
on the non-operative side of the table and lock into position.
6. Adjust table height to appropriate position to transfer patient from patient stretcher.
7. Attach the appropriate traction boot (6850-485) small, or (6850-486) large to patient's
operative foot while on stretcher.
8. Transfer patient to the table in supine position, insert well padded perineal post and
pad (6850-413) move the patient to establish firm contact with the perineal post and
secure with patient safety strap (5840-44).
9. Place operative leg traction boot into traction mount on the operative leg spar.
10. Attach well leg holder (5855-838) to the well leg support adaptor (6875-200) using the
articulating bracket (6300-92) and secure patient's leg.
11. Install arm boards with pad (6875-292) to each side of the table and position patient's
arms out to the side, or attach arm boards with pad (6875-292) to the non operative
side of the table and optional cross arm support (5857) using clark socket (5393) to
the non-operative side of the table and secure the patient's arms.
12. Remove patient transfer board (6875-742) by unclamping it from the leg spar and
sliding it out from underneath the fully positioned patient.
13. Install the head end drape rod assembly (6875-747) at the head end of the table and
the foot end drape rod assembly (6875-750) on the appropriate leg spar at the foot
end of the table with hole on the outside. Adjust to the appropriate height and lock.
MIZUHOSI 2009
49
®
Surgery Table Set-Up
NW0508 Rev. D