6.4 Power Cord

The table is equipped with a detachable standard IEC power cord connector. The power cord
is connected to the table at the IEC power entry socket located below the on/off switch, at the
head end control panel.

6.5 Battery System

®
The 6875 hana
Surgery Table is equipped with a battery system that provides power to all
functions, and consists of two batteries (NV0801).
consumption of the 6875 hana
to 12 hours at 10% duty cycle.
This symbol indicates an external ground stud that is required for use when the
AC power cable is not connected to a protective earth ground hospital grade AC outlet
in your operating room or facility.
To protect the patient, hospital staff and the table from possible electrical hazards, an
external ground wire connection is required between the external ground stud and
protective earth ground.
Battery charge status is shown via a LED on the hand pendant labeled battery status and in
the power indicator section of the control panel.
On the hand pendant:
 If illuminated green = OK.
 If illuminated red = charge required
If a recharge is required, charging the batteries is necessary for a minimum of 3 hours. Fully
discharged batteries will require 18 hours for a complete charge. To charge the batteries,
simply plug the power cord into an appropriate hospital grade AC outlet and turn the power
switch on. It is important to plug in the table as soon as possible after the battery status LED
is illuminated red.
MIZUHOSI 2009
On/off power
switch
Power
cord
receptacle
Figure 12: Head end power supply box and control panel
®
Surgery Table, the table can be used on battery power for up
WARNING:
Due to the relatively low power
19
NW0508 Rev. D