Validated cleaning and disinfection procedure
Validated procedure
Special features
Wipe disinfection for
electrical devices with-
out sterilization
5.7
Wipe disinfection for electrical devices without sterilization
Phase
Step
I
Wipe disinfection
RT:
Room temperature
Phase I
Remove any visible residues with a disposable disinfectant wipe.
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
Observe the specified application time (1 min minimum).
5.8
Inspection, maintenance and checks
Allow the product to cool down to room temperature.
Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
Check the product for any damage, abnormal running noise, overheat-
ing or excessive vibration.
Set aside the product if it is damaged.
Keep the optical cable interfaces clean.
Be sure that the optical cable interfaces are not scratched. Rub the
optical cable interfaces with a wad of cotton.
T
t
Conc.
[°C/°F]
[min]
[%]
≥1
RT
-
Water quality
Chemical
-
Meliseptol HBV wipes 50 % Propan-1-ol
6.
Maintenance
To ensure reliable operation, the product must be maintained at least once
a year.
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
The guaranteed service life of the xenon lamp is 500 hours of operation.
Note
Maintenance must only be carried out by authorized personnel. If neces-
sary, the circuit diagrams and the service manual, which contains all the
necessary documents, are provided for this.
6.1
Replacing xenon lamp
Risk of fatal injury from electric shock!
DANGER
Danger of explosion due to incorrect handling!
DANGER
Danger of burns from a hot xenon lamp!
DANGER
Reference
Chapter Wipe disinfection for electrical devices
without sterilization
Pull out the power plug before opening the
device.
The xenon lamp should only be operated by
qualified personnel.
Always wear protective gloves and safety
glasses.
Only replace the xenon lamp after the device
has completely cooled down.