Closing the device
Note
Before putting the device cover in place, the connection between the cover
and the ground cable must be present and intact.
Pull the device cover slightly outwards, by its lower edge.
Set the cover on the device.
Make certain that no wires are caught.
Press the fast-action locks 17.
Rotate the fast-action locks 17 clockwise 90° with a screwdriver until
it snaps into place.
7.
Troubleshooting list
Malfunction
Xenon lamp and fan do not work
Xenon lamp does not come on
Fan does not work
Brightness noticeably poorer
Detection
Device not powered
Indicators not illuminated
Indicator light not illuminated even
after change of fuse
No light comes out of end of opti-
cal cable
No air comes out of right grating
The optical cable cannot be
removed
Visual check of filter
Visual check of filter
Setting back the operating time counter
The xenon light source is equipped with an integrated operating time
counter that registers the number of hours the xenon lamp has been in
operation. If the xenon lamp exceeds the maximum permissible number of
hours in operation of 500 hours, the lamp change lamp indicator 8 lights
up. After inserting the new xenon lamp, the operating time counter must
be reset.
Switch on the device.
Press on the two the lamp OK indicator 4 and change lamp indicator 8
displays for approximately 3 seconds.
This sets the counter back.
The lamp OK 4 and change lamp 8 indicators flash alternately four times.
Then the lamp OK indicator light 4 is illuminated.
Cause
The plug is not inserted in the
socket
Fuses blown
Malfunction in power supply unit
Power ON switch not activated
The xenon lamp is malfunctioning
Malfunction in power supply unit
Xenon lamp overheated
Malfunction in fan
Optical cable not completely
inserted
Thermal protection filter dirty
Malfunction in thermal protection
filter
Remedy
Insert plug in socket
Replace fuses, see Fuse replace-
ment
Have xenon light source repaired
by manufacturer
Activate power ON switch
Replace the xenon lamp
Have xenon light source repaired
by manufacturer
Wait a few moments until the
xenon lamp is ignited
Have xenon light source repaired
by manufacturer
Use the correct optical cable con-
nection
Clean the thermal protection filter
Have xenon light source repaired
by manufacturer