Emc Declaration - Dräger PressurePod Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 32
English / English
USEnglish / English US
Weight
PressurePod, including
<700 g (1.5 lb)
USB cable and rail
clamp
USB cable
Length
>140 cm (55.1 in)
Measured values
Paw/Paux1, Pes/Paux2,
Pga/Paux3 Pressure measurement
Range
±90 mbar or cmH
Accuracy
±2 mbar or 3 % of the mea-
sured value
Classifications
Liquid ingress and solid
IP 32 according to IEC
particle protection
60529
Tested when ready
for operation (metal
housing with plastic
cover)
Use of latex
The product is not made
with natural rubber latex
GTIN
04048675625612
GMDN code
31278
Classification
Class IIa
Medical Device Europe
UMDNS code
14-119
Universal Medical De-
vice Nomenclature Sys-
tem –
Nomenclature for medi-
cal devices
20
Device interfaces
USB plugs
Connect only to compat-
ible devices with USB
2.0 interface whose lists
of accessories include
the PressurePod as an
accessory.
Pressure measuring ports
Paux1/Paw port
O
2
Paux2/Pes port
Paux3/Pga port
Max. permissible input
pressure
Mounting

EMC declaration

General information
This device was tested for electromagnetic compati-
bility. Third-party accessories may only be used if
they do not compromise the electromagnetic com-
patibility. The use of non-compliant accessories may
result in increased electromagnetic emissions or de-
creased electromagnetic immunity of the device.
This device may be used in the direct vicinity of other
devices only if Dräger has approved this device ar-
rangement. If no approval has been given by Dräger,
it must be ensured that this device functions correctly
in the desired arrangement before use. The instruc-
tions for use for the other devices must be followed.
Electromagnetic environment
This device may only be used in environments spec-
ified in section "Environments of use" on page 14.
Emissions
Radiated emissions Class A, group 1 (30 MHz to
Conducted emis-
sions
Luer Lock male
NOTE
Luer Lock male
The emissions characteristics of this equipment
Luer Lock male
make it suitable for use in industrial areas and hos-
pitals (CISPR 11 class A). If it is used in a residen-
±200 mbar or cmH
O
2
tial environment (for which CISPR 11 class B is
normally required), this equipment might not offer
Compatible with:
adequate protection to radio-frequency communi-
DIN standard rail
cation services. The user might need to take miti-
SMP Fairfield rail
gation measures, such as relocating or re-orienting
the equipment.
Immunity against
Electrostatic discharge
(ESD) (IEC 61000-4-2)
Magnetic fields at mains
frequency
(IEC 61000-4-8)
Radiated high-frequen-
cy disturbances
(IEC 61000-4-3)
Conducted high-fre-
quency disturbances
(IEC 61000-4-6)
Compliance
1 GHz)
Class A, group 1 (150 kHz to
30 MHz)
Test level and required
electromagnetic envi-
ronment
Contact discharge: ±8 kV
Air discharge: ±15 kV
50 Hz: 30 A/m
80 MHz to 2.7 GHz:
3 V/m
150 kHz to 80 MHz: 3 V,
ISM bands: 6 V
Instructions for use PressurePod

Publicidad

Tabla de contenido
loading

Tabla de contenido