Service; Inspection; Disposal; Technical Data - Dräger PressurePod Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 32

Service

Safety information
WARNING
Risk due to inappropriately reprocessed prod-
ucts
The product may be contaminated with infec-
tious agents.
Before service is performed and before the
product is sent back for repair, reprocess the
product in accordance with the chapter "Re-
processing.
WARNING
Risk if service is not performed regularly
Wear and material fatigue of the components
may lead to device failure and malfunctions.
Perform service at the specified intervals.
WARNING
Risk when the housing is being opened
Under the housing, there are live electrical
components, which may cause device failure
and malfunctions if they are not handled cor-
rectly.
The housing may only be opened by those user
groups that are assigned to that particular
measure.
Definition of service terminology
Concept
Definition
Service
All measures (inspection, mainte-
nance, repair) intended to maintain
or restore the functional integrity of
a product
Instructions for use PressurePod
Concept
Definition

Inspection

Measures intended to determine
and assess the current state of a
product
Maintenance
Regular specified measures in-
tended to maintain the functional
integrity of a product
Repair
Measures intended to restore the
functional integrity of a product af-
ter a failure
Inspection
.
Measure
Interval
User group
Inspection
Every 2 years Service personnel
Zero calibra-
Every 2 years Service personnel
tion of the
pressure sen-
sors of the
PressurePod
Performing inspection
1
Check that the respective instructions for use are
present.
2
Check that the product is in good condition:
All labels are complete and legible
There is no visible damage

Disposal

This product must be disposed of in accordance with
national regulations.
English / English US

Technical data

All given measurement characteristics and toleranc-
es are valid at 20 °C (68 °F), 60 % relative humidity,
and 1013 hPa (760 mmHg).
Ambient conditions
During operation
Temperature
5 to 40 °C (41 to 104 °F)
Ambient pressure
620 to 1100 hPa (9.0 to
15.9 psi)
Relative humidity
5 to 95 %, without conden-
sation
Height
Up to 4000 m (13123 ft)
During storage and transportation
Temperature
20 to 60 °C (
Ambient pressure
500 to 1100 hPa (7.3 to
15.9 psi)
Relative humidity
10 to 95 %, without con-
densation
Operating characteristics
Power supply
Port
USB 2.0 compatible inter-
face for connecting a com-
patible device
Voltage
5 V ±10%
Current
<100 mA
Medical device
Protection class II
Dimensions of the housing (without rail clamp)
Height (without Luer
<115 mm (4.5 in)
Lock connector)
Width
<125 mm (4.9 in)
Depth
<65 mm (2.6 in)
4 to 160 °F)
19

Publicidad

Tabla de contenido
loading