5.6
Wipe disinfection for electrical devices without sterilization
Phase
Step
I
Wipe disinfection
RT:
Room temperature
Phase I
Remove any visible residues with a disposable disinfectant wipe.
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
Observe the specified application time (1 min minimum).
5.7
Inspection, maintenance and checks
Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
Inspect all cables, and in particular electrode cables, for any damage to
their insulation.
Set aside the product if it is damaged.
A technical safety inspection must be carried out annually, see Main-
tenance.
5.8
Storage and transport
Damage to the product caused by switch-on to
early after storage/transport at temperature below
+10 °C!
CAUTION
Allow the HF device to acclimate at room tem-
perature for about 3 hours.
Transport the product only in its original box.
Observe storage and transport conditions, see Ambient conditions.
T
t
Conc.
[°C/°F]
[min]
[%]
≥1
RT
-
Water quality
Chemical
-
Meliseptol HBV wipes 50 % Propan-1-ol
6.
Maintenance
The high-frequency surgical device GN300 is a maintenance-free product.
Note
Maintenance must only be carried out by authorized personnel. If neces-
sary, the circuit diagrams and the service manual, which contains all the
necessary documents, are provided for this.
6.1
Safety inspection
A safety inspection must be carried out annually.
The person in charge of the inspection will document the inspection find-
ings and measurement values according to the printed inspection proto-
col.
Only have the product and its accessories inspected by persons who
possess the requisite training, knowledge and experience and who are
authorized to do so.
If the maximum values specified in the enclosed inspection protocol
for the technical safety checks are exceeded: send in the device.
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.