High-frequency surgical device GN300
Fig. 9
Correction curves, Bipolar CUT (P = f(R) original cable
CL = 175pF)
Fig. 10 Correction curves, Bipolar COAG (P = f(R) original cable
CL = 175pF)
2.10 Maximum peak output voltage (U
Note
The following graphs allow the user to judge whether the HF surgical device
or its output setting is suitable for a given accessory (insulation rating).
Risk of injury to patients or users caused by inade-
quate accessories (insulation rating)!
Ensure that the accessory voltage rating speci-
WARNING
fied in the product documentation is higher
than the maximum (settings-dependent) peak
output voltage.
10
®
)
p
Fig. 11 Maximum peak output voltage (Up) of the monopolar output in
CUT mode
Fig. 12 Maximum peak output voltage (Up) of the monopolar output in
COAG mode
Fig. 13 Maximum peak output voltage (Up) of the bipolar output