Aesculap
High-frequency surgical device GN300
5.3
Preparation before cleaning
Disconnect the unit from mains power.
Remove accessories.
5.4
Preparations at the place of use
Remove any visible surgical residues to the extent possible with a
damp, lint-free cloth.
Transport the dry product in a sealed waste container for cleaning and
disinfection within 6 hours.
5.5
Cleaning/disinfection
Product-specific safety instructions for the reprocessing procedure
Risk of electric shock and fire hazard!
Unplug the device before cleaning.
DANGER
Do not use flammable or explosive cleaning or
disinfecting solutions.
Ensure that no fluids will penetrate the product.
Damage to, or destruction of the product caused by
mechanical cleaning/disinfection!
Only clean and disinfect the product manually.
CAUTION
Do not sterilize the product under any circum-
stances.
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
CAUTION
Only use cleaning/disinfecting agents approved
for surface cleaning. Follow the manufacturer's
instructions for the respective cleaning/disin-
fecting agent.
Observe specifications regarding concentration,
temperature and exposure time.
Validated cleaning and disinfection procedure
Validated procedure
Special features
Wipe disinfection for
electrical devices with-
out sterilization
20
®
Reference
Chapter Wipe disinfec-
tion for electrical devices
without sterilization