5.3
Preparation before cleaning
►
Disconnect the unit from mains power.
►
Remove accessories.
5.4
Preparations at the place of use
►
Remove any visible surgical residues to the extent possible with a
damp, lint-free cloth.
►
Transport the dry product in a sealed waste container for cleaning and
disinfection within 6 hours.
5.5
Cleaning/disinfection
Product-specific safety instructions for the reprocessing procedure
Risk of electric shock and fire hazard!
►
Unplug the device before cleaning.
DANGER
►
Do not use flammable or explosive cleaning or
disinfecting solutions.
►
Ensure that no fluids will penetrate the product.
Damage to, or destruction of the product caused by
mechanical cleaning/disinfection!
►
Only clean and disinfect the product manually.
CAUTION
►
Do not sterilize the product under any circum-
stances.
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
CAUTION
►
Only use cleaning/disinfecting agents approved
for surface cleaning. Follow the manufacturer's
instructions for the respective cleaning/disin-
fecting agent.
►
Observe specifications regarding concentration,
temperature and exposure time.
Validated cleaning and disinfection procedure
Validated procedure
Special features
Wipe disinfection for
–
electrical devices with-
out sterilization
20
®
Reference
Chapter Wipe disinfec-
tion for electrical devices
without sterilization