4.5.3
Metal pipe with loose flanges
1. Solder the threaded nuts M6 (1) to the pipeline and connect the ground as shown in the
illustration.
2. Use a copper wire (at least 2.5 mm² (14 AWG)) to establish the ground connection between
the sensor (2) and an appropriate grounding point.
Fig. 22: Metal pipe, without liner (example)
IMPORTANT (NOTE)
•
Earthing is illustrated using the example of the dual-compartment transmitter housing; in
the case of transmitters with single-compartment housing, earthing is to be performed as
shown
CI/FEX300/FEX500-X1
Flange design
FEX300, FEX500
Mounting
Wafer-type design