Commissioning
6.5.2

Parameterization via the "Commissioning" menu function

The device can be factory parameterized to customer specifications upon request.
If no customer information is available, the device is delivered with factory settings.
The setting of the most current parameters is summarized in the "Commissioning" menu. This
menu provides the quickest way to configure the device.
The "Commissioning" menu allows you to select the language, the physical unit for flow rate, the
measuring range, the totalizer unit, the pulse / frequency mode, the pulse per unit, the pulse
width, damping, and the status of the current output during an alarm (Iout for alarm, Iout: Low
Alarm, Iout: High Alarm).
For detailed descriptions of these menus and parameters, see the chapter on the "Parameter
overview".
The following section describes parameterization via the "Easy Setup" menu function.
Boiler 9
Q
Σ+
Σ-
Read only
Standard
Advanced mode
Back
**********
RSTUVWXYZ012345
Next
Exit
54 - EN
0.00
l/s
0000.00 m³
0000.00 m³
Access levels
Select
Enter password
OK
Menu
Easy Setup
Select
4. Use
to switch to the configuration level.
5. Use
or
6. Confirm the selection with
7. Use
to confirm the password. A password is not
available as factory default; you can continue without
entering a password.
8. Use
or
9. Confirm the selection with
FEX300, FEX500
to select "Standard".
.
to select "Commissioning".
.
CI/FEX300/FEX500-X1