Installing canister

1.
Ensure device is off.
2. Remove paper tape around canister tubing to
release tubing to full length.
3. Open both orange clips on sides of canister.
4. Align canister so that viewing window is facing
forward.
5. Push canister gently over device inlet port.
6. Engage both canister clips. Clips will click when
properly engaged.
7. Following canister installation, to verify functional
complete blockage alarm, turn device on and insert
tethered cap of canister tubing into connector
to simulate a blockage. A functional device will
activate a complete blockage alarm within 5
minutes.
In the case that complete blockage alarm does not
activate, check canister installation and contact
your Smith & Nephew authorized representative.
*smith&nephew
RENASYS™ GO
Negative Pressure Wound
*smith&nephew
Therapy
RENASYS™ GO
Negative Pressure Wound
Therapy

Removing or changing canister

1.
Hold orange quick click connector above wound to
help ensure exudate does not leak from tubing.
2. Turn device off.
3. Disconnect canister tubing from Soft Port.
4. Insert tethered cap into both sides of orange quick
click connector.
5. Release orange canister clips on sides of canister
and gently pull away from device.
Disposal of used canisters should follow
facility protocols or local ordinances
relating to handling of potentially
infected or biohazardous materials.
Connect RENASYS™ Soft Port dressing
and canister
Connect Soft Port to canister tubing by pushing orange
quick click connectors together. An audible click
indicates connection is secure.
Disconnect RENASYS Soft Port dressing
and canister
*smith&nephew
RENASYS™ GO
Negative Pressure Wound
*smith&nephew
Therapy
RENASYS™ GO
Negative Pressure Wound
Therapy
1.
2. Turn device off.
3. Disconnect canister tubing from Soft Port.
4. Insert tethered cap into both sides of orange quick
12
Note: Do not cover aeration disc.
Hold orange quick click connector above wound to
help ensure exudate does not leak from tubing.
click connector.