Maintenance

Inspect device for visible signs of damage before each
use. If device has been dropped or shows signs of
damage, do not use. Return damaged device to your
Smith & Nephew authorized representative.
There are no serviceable parts in the device. Do not
attempt to open device housing. Contact your
Smith & Nephew authorized representative or
distributor if service is required.
Cleaning
Adherence to facility directives concerning hygiene
is of prime importance. The instructions supplied
with all cleaning agents as well as sterilization and/or
disinfection units must be followed.
Cleaning of the RENASYS™ GO outer housing should be
done in accordance with the guidelines below:
Wipe down device surface with a damp soft cloth.
Use a mild cleaning agent, disinfectant, or an
antimicrobial agent and ensure its compatibility
with plastics. Follow manufacturer's guidelines for
use of cleaning or disinfectant agents.
Dampen another soft cloth with clean water and
wipe down all surfaces to remove any excess
solution.
Dry with a separate soft cloth.
Do not use solvents or abrasives.
Do not immerse any part of RENASYS GO in
fluid, directly spray any cleaner onto device or
use a wet cloth that is dripping fluid. No fluids
should be allowed to enter device. If any liquids
penetrate device, contact your local distributor.

Device replaceable components

RENASYS GO replaceable components including
the Inlet O-ring, odor filter and rubber feet may
be purchased through Smith & Nephew. Contact
your Smith & Nephew authorized representative or
distributor for more information.
Inlet O-ring (REF 66800603)
To ensure the canister seal is secure, the device inlet
port O-ring should be replaced after each patient use
as the O-ring can become worn over time, potentially
creating a source of air leak that may impact alarm
functionality.
1.
To replace the device inlet port O-ring, the
existing O-ring must be carefully removed using
a plastic removal tool, ensuring that the inlet
connector is not damaged during removal.
2.
Clean the device inlet port and carefully slide new
inlet port O-ring onto device inlet port; checking
that any rolling of the O-ring is corrected once it is
located in the groove of the inlet port.
DO NOT LUBRICATE THE O-RING!
NOTE: Smith & Nephew does not offer a plastic
removal tool to be used with the RENASYS GO device.
Plastic removal tools may be purchased from vendors
such as Golem Gear (SKU GG_Yellow Pick).
Odor filter (REF 66800061)
To minimize unpleasant odors from RENASYS GO
devices, it is recommended that the device odor filter
be replaced between each patient use.
Note: The presence of device odor does not impact
the functionality of the device.
1.
To change or verify the presence of the odor
filter, turn the device over and lay it on a soft
surface, to avoid damage.
2.
Remove the Posidriv
and open the flap to reveal the odor filter.
3.
Remove the two odor filter elements and replace
with 2 new odor filter elements.
4.
Close the door flap and insert screw, tightening
only until the door flap is flush with the case.
Precaution: TAKE CARE NOT TO OVER-TIGHTEN
SCREW!
Rubber feet (REF 66800611)
If a rubber foot has become detached or is
missing, it may be replaced. Use a small quantity of
cyanoacrylate adhesive to reattach the rubber foot to
the base of the device. Rubber feet can be obtained as
replacement parts.

Storage

Prior to storage for more than 2 months, ensure that the
battery is fully charged. Some battery discharge may
occur in storage. During long-term storage, recharge
the battery every 10 months.
RENASYS GO should be stored between 0–25°C
(32–77°F) for optimal battery performance, but can be
stored between -10–55°C (14–31°F) for short periods
of time.

Returning device

Prior to returning the device to your Smith & Nephew
authorized representative, device must be cleaned
according to the steps outlined under the cleaning
section of this user manual.
Device should be returned within the original packaging.