Load the autoclave to maximise drying performance. For example, stack transparent sterilization wrap
and paper packages like index cards. Follow the instructions in the section Loading the autoclave on
page 15. Use the film holder if necessary.
Activate the Additional drying function. Follow the instructions in the section Select extended
drying on page 20.

Program has ended

If the program has successfully ended, the display shows the message:
If immediate output at program end has been activated in the "Function" set-up menu Last batch no.,
the log for the completed program is issued on the activated output media after the door is opened (see
page 25, Chapter 5 – Logging).

Manual termination of a program

You can interrupt a running program in all phases. However, if you terminate a program before the start of
the drying, the load is still non-sterile. The program is considered not to have run successfully.
WARNING
Terminating a running program by switching the power switch off can lead to hot steam
escaping from the sterile filter resulting in its contamination.
Never terminate a running program by switching the autoclave off at the power switch.
DANGER!
The sterilization chamber, the door, and the items that have been sterilized are hot. After a
program has been interrupted, hot steam may also escape when the door is opened depending
on when the program was terminated.
Burning hazard due to hot steam.
Only remove the trays using the tray lifter.
Never touch the items that have been sterilized, the sterilization chamber or the inside of the
door with unprotected hands.
Manual termination during the drying phase
During the drying phase you can interrupt the program using the (S) key without the autoclave reporting a
malfunction.
You must then expect inadequate drying, particularly for wrapped items. The sterilized items must be
sufficiently dry before proper sterile storage is possible. For this reason, you should allow the program to
run through to the end of the drying phase when possible for programs with wrapped items.
In a quick program, sterilized unwrapped instruments dry after being removed from the autoclave because
of their own heat.
Universal program
run successfully
Chapter 4 – Sterilizing
Programm
Start-Stop
21