Chapter 2 – Installation
Chapter 2 – Installation
NOTE
Please carefully follow the instructions for the installation given in the service manual. All
requirements for the building installation are listed in detail in the service manual.

Electrical connections

DANGER!
Improperly made electrical connections can lead to a short circuit, fire, water damage, and/or
electric shock.
Severe injuries may result.
Please comply with the following precautionary measures when handling power cables and power plugs.
Never splice or modify the power cable.
Never bend or twist the power cable.
Always hold the power plug itself to remove it from the power plug.
Do not place heavy objects on the power cable.
Ensure that the power cable is not compressed (e.g., between doors or windows).
Do not place the power cable in front a heat source.
Do not use any nails, staplers or similar objects to fix a cable.
Should the power cable or the power plug be damaged, shut the autoclave down. The power cable or
power plug may only be replaced by authorized persons.
Failure to comply with this precaution can lead to damage to the cable or to the plug and/or to fire or
an electric shock. Severe injuries may result.

Feed water connection

For steam sterilization, the use of distilled or demineralized water, referred to as feed water, is required.
The DIN EN 13060 intends that feed water in accordance with the recommended values in Annex C is
used. The feed water is supplied either via the internal water storage tank or via a separate water
treatment unit; see
treatment unit can be found in the service manual.

Waste water connection

The waste water can either be collected in the internal water storage tank on the waste water side (left)
and manually drained or automatically drained via the one-way drain. To connect the autoclave to the
waste water, a retrofit kit can be ordered for the tank drain.
10
The electrical connection and the connections to the feed and waste water must only be
carried out by a licensed electrician or plumber respectively.
Also note the information regarding the installation and initial start-up in the service manual.
Chapter 3 – First
steps. Detailed information about the connection to a water