Checking the water quality of the feed water
Whenever the autoclave is switched on, including when a program is running, you can show the water
quality on the display.
To do so, hold down the (-) key until the message Conductivity appears on the display. The
conductivity is shown in µS/cm.
As soon as you release the (-) key, the display returns to its previous status (e.g., the initial operating
state).
Checking the preheat temperature of the chamber
If preheating is activated, the autoclave preheats the cold chamber or maintains the temperature between
two sterilization cycles. This is used to shorten the program times and to reduce the formation of
condensation in order to improve the drying results.
If you press the (-) key 2×, first briefly and then holding it down the second time, instead of the conductivity
the preheat temperature of the chamber is shown.
AIN6: Conductivity
15 µS/cm
AIN4: Temp. preheat
120 °C
Chapter 6 – Functional testing
Programm
Start-Stop
Start-Stop
Programm
33