General incidents
Incident
No reading on the display
Door will not open
Feed water consumption
too high
Poor drying results
Possible cause
No power
(item 2)
Door seal adheres to the sealing
surface.
Autoclave has been loaded
incorrectly
Autoclave is not positioned correctly.
The condensate drain is blocked.
Autoclave has been loaded
incorrectly.
Autoclave is not positioned correctly.
The condensate drain is blocked.
Chapter 10 – Malfunctions
What you can do:
Check that the power plug is
correctly inserted into the socket.
Check the mains voltage at the
socket.
If necessary, change the device
fuses on the lower front of the
autoclave (see page 8, Fig.
1/10). Follow the instructions in
the service manual under
"Change device fuses."
Switch the autoclave on, press
the "+" key to open the door, and
pull forcibly on the door to open
it.
Pay attention to the load size
(page 15, "Loading the
autoclave").
Check that the autoclave is
positioned correctly. If
necessary, unscrew the leveling
feet by a max. of 2 rotations.
Remove any instruments, filter
paper or similar that have been
dropped.
Pay attention to the load sizes
(page 15, "Loading the
autoclave"). Textiles must not
make direct contact with the
walls or floor of the chamber.
Check that the autoclave is
positioned correctly. If
necessary, unscrew the leveling
feet by a max. of 2 rotations.
Remove any instruments, filter
paper or similar that have fallen
down.
Check the chamber filter and the
filter of the condensate drain for
any blockages.
Activate pre-heating (see page
19, Select automatic preheating)
Activate additional drying (see
page20, "Additional drying").
43