Warning
Log memory full
Please carry out
maintenance
Test not successful
Leak rate: 3.2
Caution! Battery empty
Possible causes
Power supply to the printer
has been disconnected.
Printer is offline.
The internal log memory of
the device is full (max. 40
logs possible).
An output medium is
registered and in the Log
output menu the option
Immediate output no is
set.
The maintenance message
is activated because the
device has reached the
specified number of batches.
The leak rate determined in
the vacuum test is greater
than the maximum permitted
value of 1.3 mbar.
Door seal, chamber flange
soiled.
Door seal not correctly
seated.
The battery voltage monitor
for the internal battery
detected an insufficient
voltage value.
What you can do
Ensure there is power to the printer; the
red LED "P" on the NITRAprint
log printer must light up.
Set the printer to online (press the "SEL"
key on the NITRAprint;. the green "SEL"
LED must light up).
The message is shown at the start of a
program.
By pressing the (S) key repeatedly, the
message disappears and the program
starts as follows. The oldest log is deleted
in the process.
Set the autoclave to Immediate output
yes (see page26, Immediate automatic
output of logs). Delete printer memory
(see page 29,
Deleting logs in the internal log memory), if
necessary, output all saved logs first (see
page 28,
Output all saved logs). In the Log
output menu, remove the output medium
by setting the option No output
medium.
The message is shown every time a
program is started.
By repeatedly pressing the (S) key the
message disappears and the program
starts.
Keep message: Press the (S) key 2× at
the start.
Have your specialist supplier's customer
service carry out the maintenance.
NOTE! The maintenance counter will be
reset by customer service.
Check that the door seal and the chamber
flange are clean and if necessary clean
them; see page 34.
Inspect the door seal for damage and
replace it if necessary; see page 35.
Repeat the vacuum test while the
autoclave is completely cold.
Check that the door seal is correctly
seated.
Repeat the vacuum test while the
autoclave is completely cold.
The battery must be replaced by your
specialty supplier's customer service.
Chapter 10 – Malfunctions
45