Deleting logs in the internal log memory
To suppress the warning Log memory full, for example, if the option Immediate output no is set, you
can manually delete the saved logs. The following example shows how you can delete all saved logs.
Select the Function set-up menu by pressing the (+) or (-) keys at the same time.
1.
The display shows the message Function: Last batch no.
Navigate using the (+) or (-) keys until the message: Function: Log output is displayed and
2.
press the (P) key.
Navigate using the (+) or (-) keys until the following message appears on the display:
Press the (P) key to delete all logs now.
3.
To leave the submenu without deleting the logs, press the (S) key.
4.
To leave the menu after deleting the logs, press the (P) key.
5.
The display shows the message Function: Log output.
By pressing the (S) key again, you leave the menu completely and return to the initial operating state of
the display.
Reading logs correctly
Log type
File extension
Text log
Error log
Standby log
Demo log
Log header
The header of the program log includes the general information on the program that has run such as the
date, the program selected, the daily batch number, and the autoclave type.
Values for the program steps
During the program the program phases are recorded with the corresponding values for the steam
pressure, the temperature, and the time (relative to the start of the program).
Summary
The summary states whether the program was successfully completed. The values for the required
sterilization time, sterilization temperature, and pressure including their maximal deviations are also
shown.
Delete all cycles
löschen
Explanation
.PRO
Log of a successfully run program
.STR
Log for a program that failed to run successfully
.STB
Log for errors when in standby
.DEM
Log for a simulated program run.
No sterilization has actually been done.
Start-Stop
Programm
Chapter 5 – Logging