Safety Instructions
When operating the autoclave, please observe the following safety instructions as well as those contained
in subsequent chapters.
Only use the device for the purpose stated in the instruction manual.
Do not use this autoclave to sterilize any liquids.
Power cables and power plug
Never damage or change power cables or power plugs.
Never operate the autoclave if the power cable or power plug are damaged.
Never pull on the power cable to remove the power plug from the socket. Always hold the power
plug itself.
Setting up, installing, starting up
Ensure the setting up, installation and initial start-up of the autoclave is only done by persons
authorized by Sirona.
The electrical connection and the connections to the feed and waste water must only be carried
out by a skilled technician.
In accordance with current VDE specifications, the autoclave is unsuitable for operation in areas
exposed to the danger of explosion.
The autoclave is conceived for use outside the patient environment. The device should be located
a minimum of 1.5 m radius away from the treatment area.
Observe all the information contained in the service handbook for the initial start-up.
DANGER!
Failure to comply with the set-up conditions can result in malfunctions or damage to the
autoclave and/or human injury.
Preparation and sterilization
Follow the instructions of the manufacturer of the textiles and instruments for preparing and
sterilizing the textiles and instruments.
Follow the relevant standards and directives for the preparation and sterilization of textiles and
instruments, e.g., from Robert Koch Institute and the German Association for Supply of Sterile
Goods (DGSV).
Only use wrapping materials and systems that are suitable for steam sterilization according to the
manufacturer.
Program terminated
Note that when opening the door after a program is terminated, depending on when the program
was interrupted, hot steam may escape from the sterilization chamber.
Depending on when the program was terminated, the load may not be sterile. Follow the clear
instructions on the display of the autoclave. If necessary, sterilize the affected load again after
rewrapping the items.
Removing the sterilized load
Never open the door with force.
Use a tray lifter to remove the trays. Never touch the sterilized items, the sterilization chamber or
the door with unprotected hands. These parts are hot.
Check the wrapping of the sterilized items for any damage when removing them from the
autoclave. If the wrapping is damaged, rewrap the item to be sterilized and sterilize it again.
Safety Instructions
5