Guidance and manufacturer's declaration – electromagnetic immunity
The Precision Montage MRI Spinal Cord Stimulator System is intended for use in the electromagnetic environment specified below. The customer or
the user of the Precision Montage MRI Spinal Cord Stimulator System should assure that it is used in such an environment.
Immunity Test
IEC 60601
Compliance Level
Test Level
Conducted RF
3 Vrms
3 Vrms
IEC 61000-4-6
150 kHz to
150 kHz to 80 MHz
80 MHz
Outside ISM and
amateur radio
bands
6 Vrms
150 kHz to 80 MHz
Inside ISM and
amateur radio
bands
Precision montage™ mri system information for Patients
91053248-04 reV a
18 of 679
Electromagnetic Environment - guidance
Portable and mobile RF communications equipment should be used no closer to any
part of the Precision Montage MRI Spinal Cord Stimulator System, including cables, than
the recommended separation distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d=1.2√P
a
d=0.58√P
a