Boston Scientific PRECISION MONTAGE MRI Información Para El Pacientes página 22

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 35
information for Patients
a The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz;
26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz;
3.5 MHz to 4.0 MHz; 5.3 MHz to 5.4 MHz; 7.0 MHz to 7.3 MHz; 10.1 MHz to 10.15 MHz; 14 MHz to 14.2 MHz; 18.07 MHz to 18.17 MHz; 21.0 MHz to
21.4 MHz; 24.89 MHz to 24.99 MHz; 28.0 MHz to 29.7 MHz; and 50.0 MHz to 54.0 MHz.
b Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Precision Montage
MRI Spinal Cord Stimulator System is used exceeds the applicable RF compliance level above, the Precision Montage MRI Spinal Cord Stimulator
System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating the Precision Montage MRI Spinal Cord Stimulator System.
c Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Precision montage™ mri system information for Patients
91053248-04 reV a
20 of 679

Publicidad

Tabla de contenido
loading

Tabla de contenido