Boston Scientific PRECISION MONTAGE MRI Información Para El Pacientes página 21

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 35
Radiated RF
3 V/m
10 V/m
IEC 61000-4-3
80 MHz to
80 MHz to 2,5 GHz
2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
d=0.35 √P 80 MHz to 800 MHz
d=0.70 √P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
should be less than the compliance level in each frequency range.
b
may occur in the vicinity of equipment marked with the symbol shown below:
safety information
Interference
c
Precision montage™ mri system information for Patients
91053248-04 reV a
19 of 679

Publicidad

Tabla de contenido
loading

Tabla de contenido