Recommended separation distances between portable and mobile RF communications equipment and the Precision Montage MRI Spinal
The Precision Montage MRI Spinal Cord Stimulator System is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Precision Montage MRI Spinal Cord Stimulator System can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Precision Montage MRI
Spinal Cord Stimulator System as recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
RF emissions
150 kHz to 80 MHz
CISPR 11
d = 1.2 √P
W
Outside ISM and Amateur
radio bands
0.01
0.12
0.1
0.37
1
1.2
10
3.7
100
12
Cord Stimulator System
Separation distance according to frequency of transmitter (meters)
150 kHz to 80 MHz
d = 0.58 √P
Inside ISM and Amateur
radio bands
0.06
0.18
0.58
1.8
5.8
80 MHz to 800 MHz
800 MHz to 80 GHz
d = 0.35 √P
d = 0.70 √P
0.04
0.11
0.35
1.1
3.5
Precision montage™ mri system information for Patients
91053248-04 reV a
safety information
0.07
0.22
0.70
2.2