To apply the DBS Lead Stop, push the DBS Lead to the center of the Lead Stop and then
tighten the Screw. This will ensure that the DBS Lead will be inserted to the proper depth.
Note: Make sure the Lead Stop will not slide on the DBS Lead when engaged.
6. Insert the DBS Lead, with the stylet in place, into the cannula.
7. Insert the DBS Lead and the cannula into the cannula guide on the microdrive.
8. Attach the DBS Lead to the microdrive.
9. Check that the locking lever on the OR Cable connector is in the open (0) position.
10. Slide the proximal end of the DBS Lead, with stylet, into the open port on the
OR Cable connector.
Note: The descriptors "proximal" and "distal" use the Stimulator as the reference
throughout this manual.
11. Push the end of the DBS Lead into the port until it stops. Hold the DBS Lead in place
while sliding the locking lever to the locked (1) position.
12. Secure the OR Cable connector to the microdrive.
Note: Make sure the stylet is inside the DBS Lead before advancing the Lead to the
desired target.
13. Slowly advance the DBS Lead to the desired target using the microdrive.
locked
DBS Lead Implantation
Vercise™ DBS Physician Manual