Clinician Options
Clinicians and physicians can access additional options in the Remote Control to set up
the Vercise™ DBS System. These options are not available to patients.
Patient‑restricted clinician screens provide access to:
Remote Control and Stimulator clearing and relinking
Communication with the clinician programmer
Contact impedance monitoring
To access Clinician Options menu:
P
1. Press the
and
Clinician Options menu will appear as shown below.
2. Press the ▲ or ▼ button to navigate the Clinician Options menu.
Each option is discussed in sequence below.
Selection #1 – Clinician Programmer (CP) Mode
1. From the Enter Clinician Mode menu, press the
Control for communication with the Clinician's Programmer (CP).
Note: The Remote Control will remain CP Ready for 15 minutes.
2. Place the Remote Control and the IR Interface in the IR
Interface Holder with their communication ports facing.
3. Plug the serial end of the USB to Serial Adapter into
Interface Holder.
4. Plug the USB to Serial Adapter into the USB port on the
CP.
5. Power‑on the CP.
6. Launch the Bionic Navigator™ Software and wait for the IR
Communication Established display.
buttons simultaneously for approximately three seconds. The Enter
P
button to prepare the Remote
Vercise™ DBS Physician Manual
91098825-11
The Remote Control
53 of 341