All buttons are active during IR Communication and pressing any button returns
the Remote Control to the Level screen.
Whenever the Remote Control is re-activated from sleep mode the display will
default to the Level screen.
Selection #3 – Clear Link
When the "Clear Link" option is selected from the Enter Clinician Options menu, you
will immediately be required to enter the clinician's password in order to continue.
The password for the Clear Link option is A‑B‑C.
To enter the password:
The first character is highlighted when the Enter Password screen opens. To select/
confirm any character and/or move to the next character position, press
through possible characters, use ▲or ▼.
Note: Do not share this password with patients, as they may unknowingly clear the link
between their Remote Control and Stimulator.
When clearing a link:
If the password is entered correctly, the link between the Remote Control and the
"previous" Stimulator is broken immediately, and the Remote Control displays the Not
Linked screen. The Remote Control's programs remain intact. If the password is entered
incorrectly, the process is aborted and the Remote Control will return to the Enter
Clinician Options screen.
Note: If the password is entered incorrectly during an attempt to link the Remote
Control, the process is aborted and the Remote Control returns to the Sleep
screen.