Vercise™ DBS Physician Manual
Programmer Communication
The Clinician Programmer (CP) can communicate with a Stimulator. In order to begin
a programming session, the CP and the Remote Control Infrared (IR) windows must be
aligned.
1. Turn on the CP and the Bionic Navigator™ Software will boot‑up.
2. Enter the Clinician Options menu on the Remote Control by pressing the
buttons simultaneously.
3. Select CP Mode.
4. When the Remote Control displays "CP Ready," put the
IR Interface and Remote Control into the IR Interface
Holder to keep the IR windows aligned.
Note: Orient the IR Interface with the yellow circle facing
down to maximize communication.
5. Plug the serial end of the USB to Serial Adapter into the
IR Interface.
6. Plug the USB to Serial Adapter into the USB port on the CP.
7. Position the patient within 60 cm (2 ft) of the Remote
Control to ensure a complete communication link from
the programmer to the Stimulator (see "Remote Control
Position" on page 46).
Note: A message will be displayed on the Clinician
Programmer if the communication range is not
satisfactory.
For more detailed instructions on how to use the Clinician
Programmer and Bionic Navigator Software with the Vercise™ DBS System, see the
DBS Programming Manual.
Vercise™ DBS Physician Manual
91098825-11
66 of 341