Indicator appearance
N/A
Cleaning and care
It is recommended to maintain the device in good outward condition by cleaning the surface of the device regularly.
Clean with a damp cloth and a mild soap. Dry with a cloth after cleaning.
Caution: Do not use solvents stronger than isopropyl alcohol as it might degrade the material.
Caution: Do not use compressed air or pressurized water jets to clean device.
Note: Device is not delivered in a sterilized state or intended to be sterilized.
Environmental Conditions
The device can be used in a wet or humid environment and submerged in up to 1 meter-deep water for a maximum
of 30 minutes.
It can tolerate contact with: Freshwater, perspiration, urine, and mild soaps.
It can also tolerate occasional exposure to salt water, sand, dust, and dirt. Continuous exposure is not allowed.
Caution: Risk of damage to device if exposed to chlorinated water.
Dry with a cloth after contact with fresh water or humidity.
Clean with fresh water after exposure to other liquids, chemicals, sand, dust, or dirt and dry with a cloth.
See which environmental conditions apply to the device in Table 5.
Battery performance is affected by temperature below -10°C (14°F) or over 45°C (113°F), which may result in battery
warnings or prevent device to perform as expected.
Note: Prolonged exposure to temperatures below 0°C can result in device malfunction.
Table 5 Environmental Conditions
Temperature
Relative humidity
Atmospheric pressure
MAINTENANCE
The device and the overall prosthesis should be examined by a healthcare professional. Interval should be
determined based on patient activity.
Audio Feedback
3 beeps
N/A
10 beeps
2 ascending tones
2 descending tones
N/A
Use
-10°C to 45°C
(14°F to 113°F)
0% to 100%,
condensing
700 hPa to 1060 hPa
Device Status
Low Battery
Battery charge level below 15%
Error
Restart required, return to Össur for servicing if error persists
Critical Error
Device non-operational, return to Össur for servicing
Connecting
Connecting to Össur Logic app
Disconnecting
Disconnecting from Össur Logic app
Connected
Device is ON and operational and connected to Össur Logic app
Charging
10°C to 40°C
(50°F to 105°F)
0% to 100%, relative
humidity (non-condensing)
700 hPa to 1060 hPa
Shipping
Extended storage
-25°C to 70°C
0°C to 45°C
(-13°F to 158°F)
(32°F to 113°F)
10% to 100%,
10% to 90%, non-
condensing
condensing
700 hPa to 1060 hPa
700 hPa to 1060 hPa