The device may be susceptible to electromagnetic interference from portable and mobile RF communications
equipment such as mobile (cellular) telephones or other equipment, even if that other equipment complies with
CISPR EMISSION requirements.
Table 8 - Electromagnetic Emissions Compliance Information
Emissions test
Conducted and radiated RF emissions
Harmonics emissions
Voltage fluctuations / Flicker emissions
Table 9 - Electromagnetic Immunity Compliance Information
Immunity Test
Electrostatic discharge
Electrical fast transient/burst
Surge
Voltage dips
Voltage interruptions
Power frequency (50/60 Hz)
magnetic field
Conducted disturbances
induced by RF fields
Radiated RF EM fields
Table 10 - RF wireless communications Compliance Information
Test frequency (MHz)
385
450
710
745
780
810
870
930
1720
1845
1970
Standard
CISPR 11
IEC 61000-3-2
IEC61000-3-3
Standard
IEC 61000-4-2
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-11
IEC 61000-4-11
IEC 61000-4-8
IEC 61000-4-6
IEC 61000-4-3
Band (MHz)
Service
380-390
TETRA 400
430-470
GMRS 460, FRS 460
704-787
LTE Band 13, 17
GSM 800/900,
800-960
TETRA 800, iDEN 820,
CDMA 850, LTE Band 5
GSM 1800, CDMA 1900,
1700-1990
GSM 1900, DECT, UMTS
LTE Band 1, 3, 4, 25
Compliance level
Group 1 – Class B
Class A
Complies
Compliance Level
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
0 % UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycle
30 A/m
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur bands
between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Modulation
Pulse modulation 18 Hz
Pulse modulation 18 Hz
Pulse modulation 217 Hz
Pulse modulation 18 Hz
Pulse modulation 217 Hz
Immunity test level (V/m)
27
28
9
28
28
13