The patient should inspect the device regularly. If unusual movement or product wear is detected in the device at
any time, the patient should be instructed to immediately discontinue use of the device and consult healthcare
professional.
Warning: Do not attempt device repairs or remove covers. Risk of injury or damage to device. Send device to Össur
for maintenance or repair.
Noise from Foot
Noise may occur if sand or debris is present in device. In that case, the healthcare professional should doff the
foot, clean it with the help of compressed air and replace the Spectra sock if it is damaged.
Note: Do not use compressed air to clean the actuator housing.
SPECIFICATIONS
Table 6 - Power Supply Specifications
Manufacturer
Model no.
Input
Output
Power supply specif ications
Table 7 - Device Specifications
Device weight
Device build height (Fig. 2)
Ankle range of motion
Heel height adjustment range
Battery pack specification
Battery pack weight
Energy to charge the battery pack
Primary packaging weight
Primary packaging materials
Note: The primary packaging is made from recyclable materials.
ELECTROMAGNETIC COMPATIBILITY
Warning: Use of this device adjacent to or stacked with other equipment should be avoided because it could result
in improper operation. If such use is necessary, this device and the other equipment should be observed to verify
that they are operating normally.
Warning: Use of accessories and cables other than those specified or provided by the manufacturer of this device
could result in increased electromagnetic emissions or decreased electromagnetic immunity of this device and
result in improper operation.
Warning: Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the device, including cables specified by
the manufacturer. Otherwise, degradation of the performance of this equipment could result.
The device is suitable for use in any environment except in flammable/explosive atmospheres, or where exposure
to high electrical and/or magnetic fields can occur (e. g., electrical transformers, high-power radio/TV transmitters,
RF surgical equipment, CT and MRI scanners).
FRIWO
FW8030M/24
100–240 VAC, 50–60 Hz, 0.6–0.3 A
24 VDC, 1.25A
1.3 kg (2.9 lbs) (Size 24 Category 2)
1.5 kg (3.3 lbs) (Size 27 Category 5)
1.6 kg (3.5 lbs) (Size 30 Category 8)
174 mm (6 7/8'') (Size 24 Category 2)
180 mm (7 1/16'') (Size 27 Category 5)
188mm (7 3/8'') (Size 30 Category 8)
33°
0-50 mm (Size 27 Category 5)
Lithium-Ion / 3500 mAh / 25.2Wh
521 g (1.15 lbs)
69.9 VAh at 230 VAC / 50 Hz
59.9 VAh at 110 VAC / 60 Hz
3.1 kg (6.8 lbs)
Polypropylene (PP 5) plastic and Polyurethane Foam