NOTE
Interactive instructions for this procedure are shown in the display, so it is simply a matter of following the instructions.
This application calibrates and adjusts the following three axes of the tool:
Target axis
en
VA collimation
Dual-axis compensator (both axes)

15.2 In-the-field calibration

NOTE
The tool should be operated carefully in order to avoid vibration.
NOTE
The in-the-field calibration procedure requires great care and each step must be carried out very exactly. Imprecise
sighting or vibration of the tool can result in incorrect calibration values, leading to subsequent measurement errors.
NOTE
In the event of uncertainty the tool should be returned to Hilti Service for checking.
1.
Set up the tool in a steady, secure position on a good tripod.
2.
Select "Configuration" from the application menu.
3.
Then select "Calibration".
4.
Start the calibration procedure or, alternatively, confirm the calibration value shown and thus cancel recalibration.
90
Cancel and return to options
menu.
Select the calibration menu and
show the values saved in the
tool.
Begin calibration process.
Confirm the calibration value
shown and return to the config-
uration menu.