Processing the skin graft
Note
Only place individual carrier plates on the insertion tray, one after the other.
For long skin grafts, sequentially insert several carrier plates.
Coiling of the skin graft! The skin graft can be
lifted off from carrier plate as soon as it has been
cut into by the cutting cylinder!
CAUTION
Press the skin graft onto the carrier plate, using
fine tweezers, directly behind the cut-in line.
If the cutting cylinder has been turned too far
already, turn back the cutting cylinder, using the
screwdriver end of the ratchet.
Position the required carrier plate 5, with the grooved surface facing
up, on insertion tray 16.
Position the skin graft on carrier plate 5.
Push carrier plate 5 between guide strips 15 under cutting cylinder 2
until the cutting cylinder engages.
Turn hand ratchet 12 anticlockwise.
The carrier plate is transported.
Keep turning hand ratchet 12 until the complete carrier plate 5 is in
position on removal tray 9.
This completes the processing of the skin graft.
Changing the cutting cylinder
Note
The cutting cylinder cannot be sharpened.
Dismounting the cutting cylinder
Cuts on hands or other body parts, caused by sharp
edges on the cutting cylinder!
Touch the cutting cylinder only at the shaft.
WARNING
Turn cutting device 4 upside down so that the four screws 11 are vis-
ible, see Fig. 1.
11
11
Fig. 1
Loosen screws 11 by hand or, if they are difficult to turn, with the
screwdriver end 13 of hand ratchet 12.
Return cutting device 4 to its upright position.
4
Open upper part 17, see Fig. 2.
14
Fig. 2
Hold down lock 8 and, at the same time, pull at shaft 10 to decouple
cutting cylinder 2, see Fig. 3.
10
Fig. 3
8
17