6.5
Cleaning/disinfection
6.5.1
Product-specific safety instructions for the reprocessing pro-
cedure
Fire hazard!
►
Do not use flammable or explosive cleaning or
disinfecting solutions.
DANGER
►
Ensure that no fluids will penetrate the product.
Damage to the product due to inappropriate clean-
ing/disinfecting agents!
►
CAUTION
Only use cleaning/disinfecting agents approved
for surface cleaning. Follow the manufacturer's
instructions for the respective cleaning/disin-
fecting agent.
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
CAUTION
►
Use cleaning and disinfecting agents according
to the manufacturer's instructions which
– be approved for plastic material and high-
grade steel,
– do not attack softeners (e.g. in silicone).
– do not cause stress cracks in plastics (e.g.,
PPSU).
►
Observe specifications regarding concentration,
temperature and exposure time.
►
Do not exceed the maximum permitted cleaning
temperature of 55 °C.
►
Dry the product for at least 10 minutes at a maximum of 120 °C.
Note
The indicated drying temperature is a guide temperature only. It must be
checked taking into account the specific conditions (e.g. load) and adjusted
if applicable.
►
Do not clean motors/handpieces by ultrasonic treatment and do not
immerse them in any liquids. To avoid the risk of corrosion/malfunc-
tioning, allow any fluid that has entered the product to drain out
immediately.
9