5.3

Safe operation

Risk of burns to patient and user due to hot motor/
hot tool!
Put down the motor/hot tool beyond reach of
WARNING
the patient.
Use a cloth to protect against burns when
changing the motor or tool.
Risk of injury and material damage due to inappro-
priate use of handpieces or motors!
Always follow the safety advice and information
WARNING
given in the instructions for use of the hand-
pieces and motors.
Do not exceed the specified maximum motor
speed.
Risk of injury and/or malfunction!
Always carry out a function check prior to using
the product.
WARNING
Danger of damage to patients' hearing from sound
of motor!
When using the HiLAN pneumatic motor in
WARNING
otology, keep as much distance from the ear as
possible.
Turn safety stop 20 on HiLAN pneumatic motors GA519 and GA520R
until the mark on safety stop 20 is aligned with the "I" mark on the
motor.
Carefully press down the foot control pedal.
The motor starts running.
Adjusting speed: Press down on the pedal to a greater or lesser extent.
To select maximum speed: Press down on the pedal as far as it will go.
Switching off the HiLAN pneumatic motor, e.g., to change a tool or
handpiece intraoperatively: Turn safety stop 20 until the mark on the
unlocking sleeve is aligned with the "0" mark on the HiLAN pneumatic
motor.
5.3.1

Operation with the hand lever

Adjust the length of hand lever 17 to corresponding handpiece 25: To
do this, pull out lever extension 16 to a greater or lesser extent.
Push slide 18 to position
Carefully press down hand lever 17.
The HiLAN pneumatic motor 1 starts to run.
Adjusting speed: Press down hand lever 17 to a greater or lesser
extent.
Maximum speed: Press down hand lever 17 as far as it will go.
Switching off the HiLAN pneumatic motor, e.g., to change a tool or
handpiece intraoperatively: Turn safety stop 20 until the mark on the
unlocking sleeve is aligned with the "0" mark on the motor.
until it engages.
7