5.3
Safe operation
Risk of burns to patient and user due to hot motor/
hot tool!
►
Put down the motor/hot tool beyond reach of
WARNING
the patient.
►
Use a cloth to protect against burns when
changing the motor or tool.
Risk of injury and material damage due to inappro-
priate use of handpieces or motors!
►
Always follow the safety advice and information
WARNING
given in the instructions for use of the hand-
pieces and motors.
►
Do not exceed the specified maximum motor
speed.
Risk of injury and/or malfunction!
►
Always carry out a function check prior to using
the product.
WARNING
Danger of damage to patients' hearing from sound
of motor!
►
When using the HiLAN pneumatic motor in
WARNING
otology, keep as much distance from the ear as
possible.
►
Turn safety stop 20 on HiLAN pneumatic motors GA519 and GA520R
until the mark on safety stop 20 is aligned with the "I" mark on the
motor.
►
Carefully press down the foot control pedal.
The motor starts running.
►
Adjusting speed: Press down on the pedal to a greater or lesser extent.
►
To select maximum speed: Press down on the pedal as far as it will go.
►
Switching off the HiLAN pneumatic motor, e.g., to change a tool or
handpiece intraoperatively: Turn safety stop 20 until the mark on the
unlocking sleeve is aligned with the "0" mark on the HiLAN pneumatic
motor.
5.3.1
Operation with the hand lever
►
Adjust the length of hand lever 17 to corresponding handpiece 25: To
do this, pull out lever extension 16 to a greater or lesser extent.
►
Push slide 18 to position
►
Carefully press down hand lever 17.
The HiLAN pneumatic motor 1 starts to run.
►
Adjusting speed: Press down hand lever 17 to a greater or lesser
extent.
►
Maximum speed: Press down hand lever 17 as far as it will go.
►
Switching off the HiLAN pneumatic motor, e.g., to change a tool or
handpiece intraoperatively: Turn safety stop 20 until the mark on the
unlocking sleeve is aligned with the "0" mark on the motor.
until it engages.
7