AB69601en
■
temperature – target temperature which the device is to achieve
in this segment (needs to be at least 2°C less than the limit value
for over-temperature protection and at least 2°C greater than the
limit value for under-temperature protection),
■
time – the duration for which the set temperature ([d hh:mm]) in
days, hours and minutes must be maintained. It is possible to
select continuous operation ∞ in the final segment.
■
ramp time – the time for reaching the set temperature ([d hh:mm])
in days, hours, minutes,
■
fan - fan speed (power) in percent (only for incubators with forced
air circulation),
■
air flap – air flap position (degree of opening)
■
fan ramp - fan speed (power) when the set temperature is
reached (only for incubators with forced air circulation)
■
air flap ramp - air flap position (degree of opening) upon reaching
the set temperature.
The active value is highlighted in blue. The item highlighted in red
means that the value is out of range and you should enter a different
value, e.g. the temperature is above or below the operating range of
the device or the protection temperature limit.
By default, the fan speed (power) is set to 100%.
Reducing the fan speed can lead to operating errors,
e.g. increased temperature fluctuations and changes
(only for incubators with forced air circulation).
Ramp time – setting a short time will not accelerate
reaching the ramp, but the ramp will be reached
in the shortest possible time depending on the set
temperature, ambient conditions and the capabilities
of the heating system in the device.
The ramp parameters are factory-set in accordance with the manu-
facturer's instructions. If it is necessary to set individual parameters
when reaching the segment temperature, activate the ramp edit field
and set your own values.
If multiple parameters are shown, the window can be
scrolled up and down.
02.2023 / Rev. --
Operation
39 / 82