Cleaning, disinfection and maintenance
AB69601en
Cleaning, disinfection and maintenance
DANGER
Risk of injury due to electric shock
Disconnect the device from the electric power
supply before performing any cleaning work!
The rechargeable battery back-up serving the
controller should also be turned off.
DANGER
A danger is posed by penetrating liquids.
There is an electric shock hazard for personnel.
WARNING
There is a risk of contamination of the microbiological
sample material.
If cleaning and/or disinfection are not sufficient after
contamination, there is a risk that microbiological
sample material will be contaminated with foreign
germs.
Be conscientious and thorough when cleaning
and disinfecting.
Follow the instructions of the cleaning agent and
disinfectant manufacturers.
Observe the range of effects of the used agent.
Observe the following rules when cleaning and disinfecting the unit:
The incubator must not be cleaned in any equipment cleaning
machines.
The incubator must not be cleaned under running water, in water
baths, or using compressed air.
The units may only be cleaned manually and using liquid disinfec-
tion agents.
To prevent any corrosion due to use of detergents and disinfec-
tants, it is essential to follow the special application instructions
provided by the manufacturers of the detergent and disinfectant.
The water temperature must be between 20 and 25°C.
Only detergents and disinfectants may be used that
are in the pH range 5-8,
do not contain caustic alkalis, peroxides, chlorine compounds,
acids or alkalis.
The outside surface of the incubator housing is
powder-coated.
02.2023 / Rev. --
63 / 82