Cleaning, disinfection and maintenance
AB69601en
Cleaning, disinfection and maintenance
DANGER
Risk of injury due to electric shock
−
Disconnect the device from the electric power
supply before performing any cleaning work!
−
The rechargeable battery back-up serving the
controller should also be turned off.
DANGER
A danger is posed by penetrating liquids.
There is an electric shock hazard for personnel.
WARNING
There is a risk of contamination of the microbiological
sample material.
If cleaning and/or disinfection are not sufficient after
contamination, there is a risk that microbiological
sample material will be contaminated with foreign
germs.
−
Be conscientious and thorough when cleaning
and disinfecting.
−
Follow the instructions of the cleaning agent and
disinfectant manufacturers.
−
Observe the range of effects of the used agent.
Observe the following rules when cleaning and disinfecting the unit:
■
The incubator must not be cleaned in any equipment cleaning
machines.
■
The incubator must not be cleaned under running water, in water
baths, or using compressed air.
■
The units may only be cleaned manually and using liquid disinfec-
tion agents.
■
To prevent any corrosion due to use of detergents and disinfec-
tants, it is essential to follow the special application instructions
provided by the manufacturers of the detergent and disinfectant.
■
The water temperature must be between 20 and 25°C.
■
Only detergents and disinfectants may be used that
–
are in the pH range 5-8,
–
do not contain caustic alkalis, peroxides, chlorine compounds,
acids or alkalis.
The outside surface of the incubator housing is
powder-coated.