Operation
6.19
Correction
6.20
INTERFACE
MODBUS TCP
60 / 82
Confirms changes
Cancels the entered changes
Go to the main menu
you can correct the temperature value shown on the display by
adding the correction value. The set correction value applies to the
whole temperature range of the device. For example, if the average
temperature displayed by the device is 40.0°C and the average tem-
perature measured by an independent, external sensor is 40.5°C, the
correction should be set to +0.5°C. The mean temperature should
be calculated over a chosen period of time, such as 30 minutes. The
available correction range is between -5°C and +5°C.
The device has been calibrated by the manufacturer
in accordance with the applicable standards. The
temperature shown on the display corresponds very
closely to the temperature near to the sensor in the
chamber. It is not necessary to utilise the user's cali-
bration to ensure correct operation of the device.
Users conduct temperature correction under their
own responsibility, and must be aware of conse-
quences of changing the manufacturer's settings. If
the user calibrates the unit, the existing calibration
certificate loses its validity.
Fig. 58: User's correction
Confirms changes
Cancels the entered changes
The device allows status monitoring using the MODBUS TCP com-
munication interface.
02.2023 / Rev. --