The first option
AB69601en
It may occur that the device fails to reach the set
temperature within the set time because the set time
duration was too short. In such cases, countdown
of the segment time will start before reaching the set
temperature. Thus, the actual time of device oper-
ating in the set temperature will be shortened.
This option is available if the number of segments is 2 or more. When
the program finishes the final segment, the device starts the program
again from the first segment. You can specify if the program should
be carried out once (Loop: 1) or multiple times (Loop: 2 to 255). To
set the program to run continuously, check the option "∞". If the time
of the final segment is set to infinite, it will only be treated as infinite in
the final cycle. In other cycles it will be treated as 0.
Example:
Loop:3
Segment 1: temp. 30°C, time 2 h
Segment 2: temp. 40°C, time 2 h
Segment 3: temp. 60°C, time "∞"
The device will run Segments 1 and 2 three times and then move
on to Segment 3, which will continue indefinitely. In incubators that
do not have a cooling system, the temperature in the chamber
decreases naturally.
The created program can be started in two ways:
Go to the main menu
Then select the program you want to activate and press the Start
button
.