Operation
Class 1.0:
Class 2.0:
Class 3.1:
62 / 82
1
1
Malfunction
2
Class 1.0
3
Class 2.0
4
Class 3.1
Set temperature 44°C.
Device continues to heat as long as there is a malfunction (oscillates
around maximum temperature).
Device stops heating and cools down to ambient temperature.
The user programs the protection temperature, and if that limit value
is exceeded, the heaters are switched off and the alarm triggers. As
long as the temperature remains above the protection temperature
limit, the alarm remains on. To lower the temperature more quickly,
the door can be opened (for example). The alarm stops as soon as
the temperature is below the protection temperature limit. To restart
the device, the user must turn the device Off and then On again.
The device no longer heats up as soon as the set excess tempera-
ture is reached. Temperature fluctuates between excess temperature
and set temperature.
The user programs the protection temperature, and if that set limit
value is exceeded (due to a fault), the heaters are switched off.
Once the temperature returns to within the allowed range, the device
resumes operation.
The set temperature in the segment cannot be higher than the over-
temperature protection minus 2°C, e.g. if the over-temperature pro-
tection limit value is 50°C, then the maximum temperature that can
be set in the segment is 48°C.
The temperature protection value for protection class 3.1 ranges:
from the set temperature of +2°C to the maximum operating
temperature of the device of +10°C.
02.2023 / Rev. --
2
4
3
AB69601en