The ICP EXPRESS Monitor will erase the previously stored zero reference value and
store a new zero reference value in the ICP EXPRESS Cable whenever a new Codman
Microsensor is zeroed, or when the user chooses to adjust the zero reference
manually. The memory element in the cable is designed to perform over one million
erase/write cycles and will retain data for several years. NOTE: A transducer zero
reference value of "–99," as displayed by the ICP EXPRESS Monitor during a manual
zeroing or zero reference verification procedure, is an indication of an ICP EXPRESS
Monitor EEPROM read failure. This may be a result of a simple random timing error by
the ICP EXPRESS Monitor and can be corrected by selecting the ADJUST REFERENCE
menu item and manually adjusting the displayed zero reference value to the correct
zero reference value. However, this error can also be an indication of a failure in the
ICP EXPRESS Cable EEPROM circuit due to deterioration associated with age or
mechanical abuse. CAUTION: Discontinue use of any ICP EXPRESS Cable that
continues to yield a "–99" transducer zero reference value (as displayed on the ICP
EXPRESS Monitor screen) during the manual zeroing or zero reference verification
procedures
CAUTION: DO NOT immerse the ICP EXPRESS Cable in any liquid, or allow liquid to
enter the plug or connectors Do not use cable if connectors become wet Erratic
results can occur
CAUTION: Verify cable integrity prior to use Do not use any cable that fails to meet
the pin-to-pin resistance or electrical leakage specifications
The ICP EXPRESS Cable integrity must be evaluated by qualified technical personnel.
The cable must be routinely inspected for pin-to-pin resistance and electrical leakage.
Pin-to-pin resistance can be measured using an ohmmeter. Resistances between pins
1–1, 2–2, 3–3, and 4–4 must be one ohm or less. See Figure 8.
Electrical leakage can be evaluated using an ohmmeter. Measure the open circuit
resistance between pins 1, 2, 3, 4 (inclusive) and all other pins accessible on the ICP
EXPRESS Monitor connector end. The open circuit resistances must be 300 MΩ or
greater.
Technical Specifications
Pressure Channel
Operating Range:
Transducer Excitation Voltage:
Transducer Sensitivity (Gain):
Input Impedance:
Frequency Response:
Zero Range:
Drift:
Pressure Display Accuracy:
Pressure Display Resolution:
Isolation:
Leakage Current:
Alarms
Alarm Parameter:
Low ( ) Limit, Default:
High ( ) Limit, Default:
Low ( ) Limit Range:
High ( ) Limit Range:
External Patient Monitor Interface
Output Signal Sensitivity:
–50 to 250 mmHg
5 VDC, ±2.6%
5 μV/V/mmHg
10 MΩ
0 to 200 Hz (–3 dB)
±100 mmHg (auto pushbutton)
<0.1 mmHg/°C (exclusive
of transducer)
±1% of reading or 1 mmHg, whichever
is larger (exclusive of transducer)
1 mmHg
Isolated from ground related
circuits by >4000 Vrms.
<10 μA at 230 VAC, 50–60 Hz
Mean ICP
0 mmHg (user adjustable)
20 mmHg (user adjustable)
–50 to +249 mmHg; 1 mmHg steps
–49 to +250 mmHg; 1 mmHg steps
5 μV/Vex/mmHg or
40 μV/Vex/mmHg
(user selected; referenced to excitation
voltage of patient monitor)
Output Signal Accuracy:
Reference Signals:
Input Impedance:
Source Impedance:
Frequency Response:
Recommended External Patient
Monitor Specifications
Excitation Voltage Range:
Excitation Voltage Frequency:
Transducer Sensitivity (Gain):
Drift:
Zero Range:
Frequency Response:
Leakage:
Isolation:
Display
Type:
Active Viewing Area:
Matrix Size:
Backlight:
Viewing Angle:
Environmental
Operating Temperature Range:
Non-Operating Temperature Range
(Storage and Transportation):
Operating/Non-Operating
Humidity Range:
Operating Atmospheric
Pressure Range:
Mechanical
Size:
Weight:
Power Requirements
Internally Powered Equipment
Mode of Operation is Continuous
Rated Input:
Supply Frequency:
Supply Voltage:
Battery Operation
Type:
Operating Time:
Charge Time:
Low Battery Alert:
Safety
The ICP EXPRESS Monitor is Class I, CF Type Medical Equipment.
The ICP EXPRESS Monitor is protected against defibrillator discharge with no
interruption to normal functionality.
The ICP EXPRESS Monitor meets or exceeds the requirements of the following
standards:
IEC 60601-1 Medical Equipment (2005), Part 1: General Requirement for Safety
IEC 60601-1-2, 2nd Edition, 2001, The Electromagnetic Compatibility Requirements
for Medical Electrical Equipment
NOTE: When connecting the ICP EXPRESS Monitor to an external patient monitor,
compliance with EN55011 is achieved only when using an approved Patient Monitor
Interface Cable which has an EMI Suppression Core attached adjacent to the 9-pin
8
EN – ENGLISH
±1 mmHg or 1%, whichever is greater
(exclusive of transducer)
0, 20, and 100 mmHg
350–500 Ω
350 Ω
0 to 200 Hz (–3 dB)
0.5 to 10.0 VDC or VAC rms
DC to 5000 Hz
5 μV/Vex/mmHg or
40 μV/Vex/mmHg
<0.1 mmHg/°C
at least ±100 mmHg
at least 10 Hz
<10 μA at 230 VAC, 50/60 Hz
Isolated from ground related
circuits by 4000 Vrms.
Liquid Crystal Display (LCD)
5.6 x 3.8 cm
128 x 64 pixels
Blue-Green Electroluminescent (EL)
±30°
5°C to 45°C
–40°C to 60°C
5% to 90% (non-condensing)
700 hPA to 1060 hPA
15.5 cm H x 8.9 cm W x 15.7 cm D
2.15 kg
220/230 VAC 0.1A
50/60 Hz
220/230 VAC
Lead Acid, rechargeable
3 hours, nominal (fully charged, new
battery)
<12 hours (for fully
discharged battery)
<15 minutes life remaining