B.7  Temp  
Temp  setup  
Unit  
Appendix  C:  Guidance  and  Manufacturer's  Declaration  of  EMC  
The  monitor  is  intended  for  use  in  the  electromagnetic  environment  specified  below.  The  customer  or  the  user  of  the  monitor  
should  assure  that  it  is  used  in  such  and  environment.  
Emission  test  
RF  emissions  
CISPR  11  
RF  emission  
CISPR  11  
Harmonic  emissions  
IEC  61000-­3-­2  
Voltage  fluctuations/  flicker  
emissions  
IEC  61000-­3-­3  
The  monitor  is  intended  for  use  in  the  electromagnetic  environment  specified  below.  The  customer  or  the  user  of  the  monitor  
should  assure  that  it  is  used  in  such  an  environment.  
Immunity  test  
Electrostatic  discharge  
(ESD)  
IEC  61000-­4-­2  
Electrical  fast  
transient/burst  
IEC  61000-­4-­4  
Surge  
IEC  61000-­4-­5  
Voltage  dips,  short  
interruptions  and  voltage  
variations  on  power  
supply  input  lines  
IEC  61000-­4-­11  
74
Factory  defaults  
℃  
Guidance  and  manufacturer's  declaration  –  electromagnetic  emissions  
-­for  all  EQUIPMENT  and  SYSTEMS  
Guidance  and  manufacture's  declaration  –  electromagnetic  emission  
Compliance  
The  monitor  uses  RF  energy  only  for  its  internal  function.  
Group  1  
Therefore,  its  RF  emissions  are  very  low  and  are  not  likely  to  
cause  any  interference  in  nearby  electronic  equipment.  
The  monitor  is  suitable  for  use  in  all  establishments  other  than  
Class  A  
domestic  and  those  directly  connected  to  the  public  low-­voltage  
power  supply  network  that  supplies  building  used  for  domestic  
purposes.  
Class  A  
Complies  
Guidance  and  manufacture's  declaration  –  electromagnetic  immunity  
IEC  60601  test  level  
Compliance  level  
±6  kV  contact  
±6  kV  contact  
±8  kV  air  
±8  kV  air  
±2  kV  for  power  supply  lines  
±  2kV  for  power  supply  
lines  
±1  kV  for  input/output  lines  
 
±1  kV  differential  mode  
±1  kV  differential  mode  
±2  kV  common  mode  
±2  kV  common  mode  
<5%  UT    
<5%  UT    
(>95%  dip  in  UT)  
(>95%  dip  in  UT)  
for  0.5  cycle  
for  0.5  cycle  
40%  UT  
40%  UT  
(60%  dip  in  UT)  
(60%  dip  in  UT)  
Electromagnetic  environment  –  guidance  
Electromagnetic  environment  -­  
guidance  
Floors  should  be  wooden,  concrete  or  
ceramic  tile.  If  the  floor  is  covered  with  
synthetic  material,  the  relative  humidity  
should  be  at  least  30%.  
Users  must  eliminate  static  in  their  
hands  before  use.  
Mains  power  quality  should  be  that  of  
a  typical  commercial  or  hospital  
environment.  
Mains  power  quality  should  be  that  of  
a  typical  commercial  or  hospital  
environment.  
Mains  power  quality  should  be  that  of  
a  typical  commercial  or  hospital  
environment.  If  the  user  of  the  monitor  
requires  continued  operation  during  
power  mains  interruptions,  it  is  
recommended  that  the  monitor  be  
powered  from  an  uninterruptible  power  
supply  or  a  battery.  
33