field  strength  in  the  location  in  which  the  monitor  is  used  exceeds  the  applicable  RF  compliance  level  above,  the  monitor  should  
be  observed  to  verify  normal  operation.  If  abnormal  performance  is  observed,  additional  measures  may  be  necessary,  such  as  
reorienting  or  relocating  the  monitor  
 b.  
Over  the  frequency  range  150  kHz  to  80  MHz,  field  strengths  should  be  less  than  3  V/m.  
Recommended  separation  distances  between  portable  and  mobile  RF  communications  equipment  and  the  
The  monitor  is  intended  for  use  in  an  electromagnetic  environment  in  which  radiated  RF  disturbances  are  controlled.  The  
customer  or  the  user  of  the  monitor  can  help  prevent  electromagnetic  interference  by  maintaining  a  minimum  distance  between  
portable  and  mobile  RF  communications  equipment  (transmitters)  and  the  monitor  as  recommended  below,  according  to  the  
maximum  output  power  of  the  communications  equipment.  
 
Rated  maximum  
output  power  of  
transmitter  
(W)  
0.01  
0.1  
1  
10  
100  
For  transmitters  rated  at  a  maximum  output  power  not  listed  above,  the  recommended  separation  distance  d  in  metres  (m)  can  
be  estimated  using  the  equation  applicable  to  the  frequency  of  the  transmitter,  where  P  is  the  maximum  output  power  rating  of  
the  transmitter  in  watts  (W)  according  to  the  transmitter  manufacturer.  
NOTE  1  
At  80  MHz  and  800  MHz,  the  separation  distance  for  the  higher  frequency  range  applies.  
NOTE  2  
These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is  affected  by  absorption  and  reflection  
from  structures,  objects  and  people.  
Warning:    
M
•  
This  product  needs  special  precautions  regarding  EMC  and  needs  to  be  installed  and  put  into  service  according  to  the  
EMC  information  provided.  This  unit  can  be  affected  by  portable  and  mobile  RF  communications  equipment.  
Do  not  use  a  mobile  phone  or  other  devices  that  emit  electromagnetic  fields  near  the  unit.  This  may  result  in  incorrect  
•  
operation  of  the  unit.
Caution:    
This  unit  has  been  thoroughly  tested  and  inspected  to  assure  proper  performance  and  operation.  
•  
•  
This  machine  should  not  be  used  adjacent  to  or  stacked  with  other  equipment.  If  such  positioning  is  necessary,  this  
machine  should  be  observed  to  verify  normal  operation  in  the  configuration  in  which  it  will  be  used.  
 
76
EQUIPMENT  or  SYSTEM  –  for  EQUIPMENT  or  SYSTEM  that  are  not  LIFE-­SUPPORTING  
Recommended  separation  distances  between  
portable  and  mobile  RF  communications  equipment  and  the  Q3  monitor  
Separation  distance  according  to  frequency  of  transmitter  
150  kHz  to  80  MHz  
=
d
2 . 1
P
 
0.12  
0.38  
1.2  
3.8  
12  
 
(m)  
80  MHz  to  800  MHz  
d
P
=
2 . 1
 
0.12  
0.38  
1.2  
3.8  
12  
35
800  MHz  to  2.5  GHz  
d
P
=
3 . 2
 
0.23  
0.73  
2.3  
7.3  
23  